Sirona ORTHOPHOS SL 2D Installation Manual page 203

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016
13.1.2.2.2
Coarse and precision adjustment
The pan adjustment menus
adjustment"
secondary diaphragm"
(precision adjustment is preset. Always try to use precision adjustment
first when adjusting the unit. In most cases, previous coarse adjustment
is not necessary.
A message window indicates that a coarse adjustment is required if
necessary.
To perform coarse adjustment, uncheck the
complete the adjustment procedure in the same way as for precision
adjustment.
The correction procedure required for coarse adjustment is identical to
that for precision adjustment. The only difference between the two modes
is the size of the image field considered. Furthermore, there are fewer
auxiliary lines in the coarse adjustment mode.
13.1 General information on unit adjustment and calibration
"Sensor adjustment"
and the Ceph adjustment menu
support coarse and precision adjustment
13 Unit adjustment and calibration
"Diaphragm
and
"Adjustment of Ceph
"Fine adjustment"
box and
203