11 Commissioning, demo mode and repacking
11.12 Service routines for commissioning and repacking
172
1. Call service routine S037.4.
 Once the service routine has been selected, the IP address of the
unit is displayed in selection field 1.
 "DEFAULT", or "STATIC" or "DYNAMIC" is displayed in selection
field 2 (see table).
2. You can display various items of network data in selection field 1 by
pressing the patient symbol keys (A, B, or C) (see table).
 The patient symbol key selected in each case lights up.
3. To change the selected parameter, first press the T key (E).
4. Now use the patient symbol keys to select the required number pad
1-4 (A-D) (see also table):
A = Number pad B1
B = Number pad B2
C = Number pad B3
D = Number pad B4
 The patient symbol key selected in each case lights up.
 The digit currently selected for changing is displayed in selection
field 1 ("Digit No. 3" in the example).
Important: The number of the digit always refers to the currently
selected number pad.
 The current value of the corresponding digit is displayed in
selection field 2 ("2" in the example).
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016