Unit adjustment and calibration

DANGER
X-rays
When performing the following tests, be sure to observe the radiation
protection regulations applicable in your country (see Operating
Instructions).
DANGER
X-rays
"Radiation" is signaled by the message "X-RAY active!", a beep, and an
X-RAY LED.
IMPORTANT
If you encounter problems with unit calibration, check whether the
required EMC conditions have been met. No other heavy-duty electric
equipment (e.g. air conditioning systems, fan motors, etc.) should be
present in the vicinity of the unit.
IMPORTANT
After each unit adjustment and calibration a constancy check must be
performed.
Tip: Move the unit to a typical working height (bite block height (A) =
approx. 1,520 mm (60")) with the Up/Down keys on the control panel
before commencing calibration.