13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
220
3. Overwrite the adjustment values for S1, S2 and S3 in the text boxes
"Sensor adjustment"
of the
For information about the shifting direction (input of +/- sign in the
menu), see the chapter entitled "Shifting direction of the exposed
image area [ → 207]".
NOTICE! Use points as decimal separators!
4. Continue with the adjustment procedure as described in the chapter
entitled "Automatic adjustment: Panoramic sensor
(standard) [ → 216]".
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
menu with the measured values.
D3632.031.01.02.02 03.2016