13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
240
A
Adjustment OK
B
Adjustment not OK
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
● The exposed diaphragm area must lie
centered and straight in the image
section as well as inside the
superimposed auxiliary lines.
● A surrounding white border must be
present.
● The maximum density must lie in the
center of the diaphragm area.
D3632.031.01.02.02 03.2016
64 95 142 D3632