S021: Service for the packing position

SA**
Function
Service for packing position
No
Initiate the packing position for the diaphragm
No
Initiate the packing position for the sensor unit
* SR=service routine, ** SA=security access
11.12.5.1
S021: Test step 3
Initiate the packing position for the diaphragm
1. Call service routine S021.3 [ → 133].
 An inactive progress indicator in selection field 1 and the
2. Touch the T key (A).
 The diaphragm is moved to the packing position.
 When the packing position is reached, "0000" is displayed in
3. Exit the service routine [ → 136].
11 Commissioning, demo mode and repacking
11.12 Service routines for commissioning and repacking
message FFFF in selection field 2 signal that the system is ready
to move the diaphragm.
During this time, a progress indicator is displayed.
selection field 2.