13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
222
The needle phantom is
1. In the structure tree, under
adjustment"
element (S030.2).
"Diaphragm adjustment"
 The
area. The menu supports precision adjustment and coarse
adjustment (precision adjustment is pre-selected). Perform a
precision adjustment first. In most cases, previous coarse
adjustment is not necessary. [ → 203]
2. Establish receptivity [ → 209].
3. Create an exposure (60 kV / 3 mA; 0.3 s) [ → 209].
 The adjustment values Sx and Sy are calculated automatically
from the exposure and entered in the text boxes of the
"Diaphragm adjustment"
 The exposure is displayed in the exposure window.
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
not
in the bite block holder on the unit.
"Pan"
"Diaphragm
, click on the
menu is displayed in the action
menu.
D3632.031.01.02.02 03.2016
64 95 142 D3632