Sirona ORTHOPHOS SL 2D Installation Manual page 217

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016
13.2 Adjustment and calibration via the calibration menu
A
Adjustment OK
B
Adjustment not OK
5. Repeat the procedure starting at Step 3 until the values in the menu's
text boxes are equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT: If automatic adjustment fails to produce values equal
to zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 219].
13 Unit adjustment and calibration
The three needle images must lie in the
center and inside the auxiliary lines.
217