13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
A
B
258
3. Fit the adjusting caps (A+B) to the ear plugs and secure them with
adhesive tape.
Black adjusting cap: on the outside (sensor side)
Transparent adjusting cap (B): on the inside (tube assembly side)
4. Establish [ → 209] receptivity.
5. Create an exposure (80 kV / 14 mA; 14.9 s) [ → 209].
 The exposure is displayed in the exposure window.
 The lead balls in the adjusting caps appear as dots on the image.
 The two dots must be coincident.
If the two dots are not coincident,
6.
described in the section "Adjusting the ear plug alignment [ → 259]"
and repeat the scan.
7. Continue with the next stage of the adjustment procedure [ → 262].
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
adjust the earplug alignment as
64 95 142 D3632
D3632.031.01.02.02 03.2016