Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
ZP: Adjusting and saving the zero point
64 95 142 D3632
D3632.031.01.02.02 03.2016
S018: Test step 9
11.12.4.8
Calibrating the occlusal bite block
When calibrating the occlusal bite block, three positional values ZP, U
and L must be adjusted and saved:
ZP (zero point):
●
in.
U (upper):
●
and the bite block plate is deflected all the way up.
L (lower):
●
and the bite block plate is deflected all the way down.
Symbol on the control
panel
Patient symbol key 1 (G) visible
Selection field
1
➢ Call service routine S018.9.
The data type intended for the next storage function is displayed
in selection field 1.
The display of patient symbol key 1 (G) indicates whether or not
the bite block is connected (see table).
1. Take the occlusal bite block out of the bite block holder.
"ZP" (for zero point) is displayed in selection field 1.
The Memory key lights up.
2. To save the zero point "ZP", first press the Memory key (the R key
lights up) and then the R key.
The value for the zero point "ZP" is accepted.
11 Commissioning, demo mode and repacking
11.12 Service routines for commissioning and repacking
Zero point: applied whenever no bite block is plugged
Upper value; applied whenever the bite block is plugged in
Lower value; applied whenever the bite block is plugged in
Status
not visible
Display
ZP
U
S
Meaning
Occlusal bite block
connected
Occlusal bite block not
connected
Meaning
Zero point
Upper value
Lower value
157