13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
248
3. Create an exposure (80kV / 14 mA; 0.3 s) [ → 209].
 Adjustment value S1 is calculated automatically from the
exposure and entered in the text box of the
secondary diaphragm (Quickshot)"
 The exposure is displayed in the exposure window.
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
"Adjustment of Ceph
menu.
64 95 142 D3632
D3632.031.01.02.02 03.2016