13.2 Adjustment and calibration via the calibration menu
4. Repeat the procedure starting at Step 2 until the values in the menu's
text boxes are equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT: If automatic adjustment fails to produce values equal
to zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 242].
5. If adjustment is OK (the adjustment values in the text boxes are equal
to "zero"), save the values. [ → 209]
 The status indicator that prefixes the element
Ceph secondary diaphragm"
6. Continue with the next stage of the adjustment procedure [ → 253].
13 Unit adjustment and calibration
"Adjustment of
is checked.