Sirona ORTHOPHOS SL 2D Installation Manual page 260

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
Horizontal correction
Vertical correction
260
1. Loosen the screws (A) slightly (approx. 2 - 3 turns).
Tip: Do not unscrew them fully!
2. Adjust the ear plugs in the horizontal direction by turning the screw
(B) counterclockwise or clockwise. The direction of rotation for left-
arm (C) and right-arm (D) versions is shown in the drawing below.
3. Tighten the screw (A).
1. Turn the rotary table counterclockwise approx. 100 degrees until you
can see the screw (E) through the opening in the cover plate.
2. Loosen the screws (E) slightly (approx. 2 - 3 turns).
Tip: Do not unscrew them fully!
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 2d / cephOrthophos sl 3dOrthophos sl 3d / ceph

Table of Contents