13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
230
13.2.1.3.2
Manual adjustment: Pan - Symmetry
The manual adjustment procedure is similar to the one for automatic
adjustment [ → 216]. The only difference is that the default adjustment
values calculated automatically in the
overwritten by adjustment values calculated manually.
1. Start the
described in the chapter entitled "Automatic adjustment: Pan -
Symmetry [ → 227]".
2. After taking the exposure, measure the distances S1, S2 and S3 with
the toolbar measuring ruler [ → 197].
If possible, measure in the lower area of the needles since they may
be slightly bent after several uses.
Tip: For better measurement, you can also edit [ → 197] the exposure
using the toolbar functions.
3. Overwrite the adjustment values for S1, S2 and S3 in the text boxes
"Symmetrie adjustment"
of the
For information about the shifting direction (input of +/- sign in the
menu), see the chapter entitled "Shifting direction of the exposed
image area [ → 207]".
NOTICE! Use points as decimal separators!
4. Continue with the adjustment procedure as described in the chapter
entitled "Automatic adjustment: Pan - Symmetry [ → 227]".
"Symmetrie adjustment"
"Symmetrie adjustment"
menu with the measured values.
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
menu are
adjustment procedure as
64 95 142 D3632