Following adjustment
64 95 142 D3632
D3632.031.01.02.02 03.2016
13.2 Adjustment and calibration via the calibration menu
3. Adjust the ear plugs in the vertical direction by turning the knurled nut
(F). The direction of rotation for left-arm (G) and right-arm (H)
versions is shown in the drawing below.
4. Tighten the screw (E) firmly.
1. Swing the ear plug holders back into the beam direction (black
adjusting cap on the outside).
2. Check the adjustment of the ear plug alignment [ → 257].