4S018: Test Step 5; 5S018: Test Step 6 - Sirona ORTHOPHOS SL 2D Installation Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

11 Commissioning, demo mode and repacking
11.12 Service routines for commissioning and repacking
152
S018: Test step 5
11.12.4.4
Setting the minimum travel height
Setting the minimum travel height is possible only for a unit height that is
below the lower correction switch level (< position value of 1500)!
1. In user mode, move the unit to the required minimum travel height by
pressing the Up/Down keys (A).
2. Call service routine S018.5.
 Once the service routine has been selected, the current height
position is displayed in selection field 1.
 The Memory key (B) lights up.
3. To save the minimum travel height, touch the Memory key (B) (R key
(C) lights up) followed by the R key (C).
4. Exit the service routine.
The limitation of the minimum travel height is purely software based. The
lower limit switch is not mechanically adapted to the new minimum travel
height!
11.12.4.5
S018: Test step 6
Undoing the minimum travel height setting
1. Call service routine S018.6.
 Once the service routine has been selected, the current height
position is displayed in selection field 1.
 The Memory key (B) lights up.
2. To increase the limit for the minimum travel height, first touch the
Memory key (B) (R key (C) lights up) followed by the R key (C).
3. Exit the service routine.
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 2d / cephOrthophos sl 3dOrthophos sl 3d / ceph

Table of Contents