Checking The Radiation Time - Sirona ORTHOPHOS SL 2D Installation Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
Preparing the measurement
A
Performing measurements
Analyzing measurements
Concluding the measurement
64 95 142 D3632
D3632.031.01.02.02 03.2016
12.6

Checking the radiation time

1. Attach the Mult-O-Meter sensor in the middle of the sensor unit (DCS-
side).
2. Set the main switch (A) to I (see also Operating Instructions).
3. Wait for approx. 1 minute.
4. Press the R key.
 The unit moves to its starting position.
5. Call the Service menu and the Service routine S002.3 (see Service
Manual).
6. Use the arrow keys (A) in
60 kV/8 mA.
7. Use the arrow keys (A) in
0.5 s.
➢ Initiate the radiation. Hold the release button pressed until the set
radiation time has expired.
➢ Read the radiation time on the Mult-O-Meter.
 The value for the radiation time displayed on the Mult-O-Meter
must correspond to the radiation time of 0.5s selected in the
service routine. The permissible tolerance is ± 10 %.
 If the measured radiation time does
tolerance, replace the
 If the measured radiation time falls within the permissible
tolerance, finalize the measurement.
1. Exit the service routine.
2. Switch the unit on via switch (A) (see also Operating Instructions).
12 Startup, measurements and tests (for USA/Canada only)
12.6 Checking the radiation time
selection field 1
to select the kV/mA level
selection field 2
to select the radiation time
not
tube assembly
(see Service Manual).
fall within the permissible
183