12 Startup, measurements and tests (for USA/Canada only)
12.5 Checking the tube voltage
Preparing the measurement
A
Performing measurements
Analyzing measurements
Concluding the measurement
182
12.5

Checking the tube voltage

1. Attach the Mult-O-Meter sensor in the middle of the sensor unit (DCS-
side).
2. Set the main switch (A) to I (see also Operating Instructions).
3. Wait for approx. 1 minute.
4. Press the R key.
 The unit moves to its starting position.
1. Select the P1 program (see Operating Instructions).
2. Set the kV/mA level 62 kV/8 mA (see Operating Instructions).
3. Make ready for exposure in SIDEXIS 4 (see Operating Instructions).
4. CAUTION! Activating the release button triggers X-rays.
Start the exposure by pressing the release button. Hold down the
release button until image acquisition is completed and the acoustic
signal that indicates the end of the exposure can be heard.
➢ Read the voltage values on the display of the Mult-O-Meter.
IMPORTANT
The measured tube voltage must correspond with the tube voltage set
of 62kV. The permissible tolerance is ± 10 %.
 If the measured values
range
, replace the
 If the measured values are within the permissible tolerance
range, finalize the measurement.
➢ Switch the unit on via switch (A) (see also Operating Instructions).
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
do not fall within the permissible tolerance
tube assembly
(see Service Manual).
D3632.031.01.02.02 03.2016