Sirona ORTHOPHOS SL 2D Installation Manual page 255

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016
13.2 Adjustment and calibration via the calibration menu
A
Adjustment OK
B
Adjustment not OK
 The exposure is displayed in the exposure window.
4. Repeat the procedure starting at Step 2 until the value in the menu's
text box is equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT: If automatic adjustment fails to produce values equal
to zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 256].
5. If adjustment is OK (the adjustment values in the text boxes are equal
to "zero"), save the values. [ → 209]
 The status indicator that prefixes the
beam direction"
element is checked.
6. Remove the ceph test phantom from the sensor unit cover.
7. Continue with the next stage of the adjustment procedure. [ → 257]
13 Unit adjustment and calibration
● A horizontal bar must be visible in the
center of the image. If this bar is visible,
the exposure is OK, and the two beams
on the image are within the tolerance
range of ±10 mm.
"Adjustment of Ceph main
255

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 2d / cephOrthophos sl 3dOrthophos sl 3d / ceph

Table of Contents