03.2017 New as of: Sinius / Sinius CS / Sinius TS Installation Requirements English Installation requirements...
Page 3
Dentsply Sirona Table of contents Installation Requirements Sinius / Sinius CS / Sinius TS Table of contents General information ....................Notes on the installation prerequisites ............Structure of the document ................1.2.1 Identification of the danger levels............ 1.2.2 Formats and symbols used .............
Page 4
Table of contents Dentsply Sirona Installation Requirements Sinius / Sinius CS / Sinius TS 4.1.4 Top view with options ............... 4.1.5 Distances in the treatment room with Heliodent Plus....... 4.1.6 Side view with HELIODENT Plus ............. Sinius CS dimensions, scale bar 1:20............
General information Notes on the installation prerequisites This document describes the installation prerequisites for all versions of the Sinius / Sinius CS / Sinius TS treatment center. It contains the following information: ● Required information for practice planning. ● Information for the installer and operator regarding the necessary quality of the air and water supply media.
1 General information Dentsply Sirona 1.2 Structure of the document Installation Requirements Sinius / Sinius CS / Sinius TS Structure of the document 1.2.1 Identification of the danger levels To prevent personal injury and material damage, please observe the warning and safety information provided in this document. This...
Dentsply Sirona. The user is responsible for any damage resulting from the use of non-approved accessories. If any units not approved by Dentsply Sirona are connected, they must comply with the applicable standards, e.g.: ● IEC 60950-1 or IEC 62368-1 for information technology equipment (e.g.
2 Safety information Dentsply Sirona 2.4 Power connection Installation Requirements Sinius / Sinius CS / Sinius TS Power connection WARNING Shock hazard It is essential that you switch off the power supply PRIOR to beginning the installation. There is a risk of electric shock. People can be injured or electrical components of the unit destroyed.
Dentsply Sirona 3 On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.1 Substrate, floor On-site installation Substrate, floor Unevenness The floor must be level and horizontal in accordance with DIN 18 202. In the case of an uneven floor, the steel adapter plate must be used; see Mounting plates [ →...
3 On-site installation Dentsply Sirona 3.1 Substrate, floor Installation Requirements Sinius / Sinius CS / Sinius TS 3.1.3 Stability With a concrete floor ● Use a 12 mm dia. masonry drill to drill the holes. ● The depth of the drill hole (C) in the concrete floor must be at least 80 max.
Dentsply Sirona 3 On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.2 Requirements of the supply media Requirements of the supply media Suction machine Compressed air (free from oil) The compressor must be able to draw in air in an efficient, hygienic manner.
3 On-site installation Dentsply Sirona 3.2 Requirements of the supply media Installation Requirements Sinius / Sinius CS / Sinius TS 3.2.1 Requirements of the water supply Condition of the water 3.2.1.1 The water that supplies the treatment center must meet the following requirements to avoid operating faults.
Dentsply Sirona 3 On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.2 Requirements of the supply media Connection to the public drinking water system 3.2.1.2 Treatment center isolated from the public drinking water supply The treatment center, if it is equipped with a disinfection system, complies with the requirements of EN 1717 (free discharge with an isolation distance ≥...
3 On-site installation Dentsply Sirona 3.2 Requirements of the supply media Installation Requirements Sinius / Sinius CS / Sinius TS Installation of the water supply on the building side 3.2.1.3 ● Carry out the installation in accordance with the national installation requirements (e.g. EN 1717).
Dentsply Sirona 3 On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.2 Requirements of the supply media 3.2.2 Requirements of the air supply The air supply is required to drive the turbines, cool the drill drives, for the cooling spray and to control the treatment center.
– 6 mm: Saliva ejector and surgical suction – 11 mm: Spray aspirator ● Filter mesh size 1.8 x 2.6 mm ● Sirona recommends using cannulae with a bypass in order to prevent return flow in adhered cannulas. 3.2.3.1 Typical pressure loss in the suction system Suction flow of large suction hose red ( ○...
Dentsply Sirona 3 On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.2 Requirements of the supply media 3.2.4 Requirements of the waste water installation ● The quantity of waste water from the treatment center is max. 3 l/min. – Carry out the waste water installation accordingly.
3.3.1 Installation template We recommend that you order the installation template (REF 33 15 830) from Sirona for laying the pipe ends in the installation field. If necessary, you can also prepare the template yourself based on the diagram below.
Page 19
Dentsply Sirona 3 On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.3 Underfloor installation of supply lines Water supply Pipe 10x1 mm, corner valve outlet 3/8" Compressed air supply Pipe 10x1 mm, corner valve outlet 3/8" Suction line DN40 HT-PP ISO 8283-3 (polypropylene, inner diameter approx.
3 On-site installation Dentsply Sirona 3.3 Underfloor installation of supply lines Installation Requirements Sinius / Sinius CS / Sinius TS 3.3.2 Installation of the supply lines in the termination panel ✔ An installation template is available or was prepared. 1. Check the position of the supply lines against the installation template as per the practice blueprint.
Dentsply Sirona 3 On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.4 Cleaning the air and water pipes Cleaning the air and water pipes NOTICE Chips and other foreign materials could be flushed and/or blown into the treatment center.
3 On-site installation Dentsply Sirona 3.5 Underfloor installation of the PC connections Installation Requirements Sinius / Sinius CS / Sinius TS Underfloor installation of the PC connections Depending on the prevailing local conditions, the existing cable set can be installed in the cable duct of an underfloor installation by an installer.
Page 23
Dentsply Sirona 3 On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.5 Underfloor installation of the PC connections 2. Save the accessory parts for final installation! The preparation of the connection for the underfloor installation of the PC is completed.
4 Dimensions, technical data Dentsply Sirona 4.1 Sinius dimensions, scale bar 1:20 Installation Requirements Sinius / Sinius CS / Sinius TS Dimensions, technical data Sinius dimensions, scale bar 1:20 Dentist element with travel track 4.1.1 Distances in the treatment room Recommended distances from cabinet or wall.
Dentsply Sirona 4 Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.1 Sinius dimensions, scale bar 1:20 4.1.2 Side view Height and swivel range of the lamp LEDview Plus LEDview Monitor on lamp support tube 63 22 668 D3561...
Dentsply Sirona 4 Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.1 Sinius dimensions, scale bar 1:20 4.1.4 Top view with options Tray LEDview Plus LEDview Monitor on lamp support tube Projection of tray arm Projection of lamp arm 63 22 668 D3561 D3561.021.01.09.02 03.2017...
4 Dimensions, technical data Dentsply Sirona 4.1 Sinius dimensions, scale bar 1:20 Installation Requirements Sinius / Sinius CS / Sinius TS 4.1.5 Distances in the treatment room with Heliodent Plus Recommended distances from cabinet or wall. Center of the floor cut-out/installation area...
Dentsply Sirona 4 Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.1 Sinius dimensions, scale bar 1:20 4.1.6 Side view with Heliodent Plus Recommended installation height for wall adapter: 1110 mm (43 3/4“) Cable bushing for radiation cable The radiation cable between Sinius and the wall adapter are supplied (12.5 mm outer diameter, max.
Dentsply Sirona 4 Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.2 Sinius CS dimensions, scale bar 1:20 4.2.2 Side view Height and swivel range of the lamp LEDview Plus LEDview Monitor on lamp support tube Projection of the lamp arm in Position 1 and Position 2 63 22 668 D3561 D3561.021.01.09.02 03.2017...
Dentsply Sirona 4 Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.2 Sinius CS dimensions, scale bar 1:20 4.2.4 Top view with options LEDview Plus LEDview Monitor on lamp support tube Projection of the lamp arm in Position 1 and Position 2 63 22 668 D3561 D3561.021.01.09.02 03.2017...
4 Dimensions, technical data Dentsply Sirona 4.2 Sinius CS dimensions, scale bar 1:20 Installation Requirements Sinius / Sinius CS / Sinius TS 4.2.5 Distances in the treatment room with Heliodent Plus Recommended distances from cabinet or wall. Center of the floor cut-out/installation area...
Dentsply Sirona 4 Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.2 Sinius CS dimensions, scale bar 1:20 4.2.6 Side view with Heliodent Plus Recommended installation height for wall adapter: 1110 mm (43 3/4“) Cable bushing for radiation cable The radiation cable between Sinius and the wall adapter are supplied (12.5 mm outer diameter, max.
Dentsply Sirona 4 Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.3 Sinius TS dimensions, scale bar 1:20 4.3.2 Side view Height and swivel range of the lamp LEDview Plus LEDview Monitor on lamp support tube 63 22 668 D3561...
Dentsply Sirona 4 Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.3 Sinius TS dimensions, scale bar 1:20 4.3.4 Top view with options LEDview Plus LEDview Monitor on lamp support tube Projection of the fixed dentist element support arm 63 22 668 D3561 D3561.021.01.09.02 03.2017...
4 Dimensions, technical data Dentsply Sirona 4.3 Sinius TS dimensions, scale bar 1:20 Installation Requirements Sinius / Sinius CS / Sinius TS 4.3.5 Distances in the treatment room with Heliodent Plus Recommended distances from cabinet or wall. Center of the floor cut-out/installation area...
Dentsply Sirona 4 Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.3 Sinius TS dimensions, scale bar 1:20 4.3.6 Side view with Heliodent Plus Recommended installation height for wall adapter: 1110 mm (43 3/4“) Cable bushing for radiation cable The radiation cable between Sinius and the wall adapter are supplied (12.5 mm outer diameter, max.
4 Dimensions, technical data Dentsply Sirona 4.4 Mounting plates Installation Requirements Sinius / Sinius CS / Sinius TS Mounting plates Adapter plate If the dental treatment center is being installed as a replacement for an M1 previously installed at this location, an adapter plate is available for this purpose.
Page 43
Dentsply Sirona 4 Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.4 Mounting plates Demonstration chair plate For floors which do not permit permanent connection of the unit (e.g. demo operation at a trade show, floor heating), installation on a steel demo plate is possible.
Information on planning for the practice The following file(s) for practice planning are available for download in the dealer area of the Sirona website under DOWNLOADS => CAAD files: ● PDF file with print symbol - for to-scale printing on paper or adhesive film ●...
Dentsply Sirona 4 Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.6 Technical data Technical data Model designation: Sinius / Sinius CS / Sinius TS Power connection: 100 – 240 V AC ± 10% 50/60 Hz Rated current: 2.2 A at 240 V...
Page 46
4 Dimensions, technical data Dentsply Sirona 4.6 Technical data Installation Requirements Sinius / Sinius CS / Sinius TS Foot control wireless interface The wireless foot control can be equipped with various wireless modules. A sign is attached to the underside of the foot control indicating the wireless module used.
4 Dimensions, technical data Dentsply Sirona 4.6 Technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.6.2 Weight and packaging Dimensions of the packaging Weight (with packaging and accessories/without packaging): Sinius dentist element, assistant element: 126 cm x 72 cm x 75 cm...
Dentsply Sirona 4 Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.7 Standards/Approvals Standards/Approvals ® The treatment center Sinius complies with the following standards, among others: ● IEC 60601-1 (electrical and mechanical safety plus software reliability) ● IEC 60601-1-2 (electromagnetic compatibility) ●...
Page 50
4 Dimensions, technical data Dentsply Sirona 4.7 Standards/Approvals Installation Requirements Sinius / Sinius CS / Sinius TS When equipped with a disinfection system, the treatment center complies with the technical rules and requirements on safety and hygiene for connection to the public drinking water supply. The unit is certified according to the requirements of the DVGW (German Gas and Water Association).
EMC aspects. Sinius / Sinius CS / Sinius TS complies with the requirements for electromagnetic compatibility (EMC) according to IEC 60601-1-2:2007 Sinius / Sinius CS / Sinius TS is hereinafter referred to as "UNIT". Accessories Connecting the PC The required interface cables can be ordered from Sirona.
5 Electromagnetic compatibility Dentsply Sirona 5.2 Electromagnetic emission Installation Requirements Sinius / Sinius CS / Sinius TS Accessories for EMC measurement The EMC measurements were performed with the following PCs: PC as peripheral device for checking Fujitsu Siemens Pentium 4 Fujitsu Siemens ESPRIMO...
Dentsply Sirona 5 Electromagnetic compatibility Installation Requirements Sinius / Sinius CS / Sinius TS 5.3 Interference immunity Interference immunity The UNIT is intended for operation in the electromagnetic environment specified below. The customer or user of the UNIT should make sure that it is used in such an environment.
5 Electromagnetic compatibility Dentsply Sirona 5.4 Working clearances Installation Requirements Sinius / Sinius CS / Sinius TS Working clearances Recommended working clearances The UNIT is intended for operation in an electromagnetic environment, between portable and mobile RF where radiated RF interference is checked. The customer or the user of...
Dentsply Sirona 5 Electromagnetic compatibility Installation Requirements Sinius / Sinius CS / Sinius TS 5.5 Foot control wireless interface Foot control wireless interface Insofar as the treatment center is equipped with a foot control, one wireless module each must be installed in the foot control and in the base of the chair of the treatment center.
6 Checklist Dentsply Sirona 6.1 Installation site Installation Requirements Sinius / Sinius CS / Sinius TS Checklist Installation site We recommend performing an inspection of the circumstances on location at least 4 weeks prior to installation. The checklist should help you when doing this.
6 Checklist Dentsply Sirona 6.3 IT hardware Installation Requirements Sinius / Sinius CS / Sinius TS IT hardware Mode of operation ☑ ● Stand-alone solution (mini PC required in the base of the chair) ☐ ● PC in treatment room ☐ PC system requirements: ☑...