13 Unit adjustment and calibration
13.1 General information on unit adjustment and calibration
210
13.1.6
Test phantom for adjustment and calibration
Needle phantom for panoramic adjustment
13.1.6.1
NOTICE
Risk of damage to unit
Make absolutely sure that the needle phantom is inserted into the bite
block holder of the unit in such a way that the front needle (B) of the
phantom points
figure). Otherwise, this can result in a collision between the sensor and
the needle phantom and thereby damage the unit.
"Sensor adjustment"
For
adjustments, the needle phantom (A) must be inserted into the pan bite
block holder of the unit. For
adjustment, the needle phantom must be taken out of the bite block
holder.
Installation Manual ORTHOPHOS SL
away from the unit in the direction of the patient
"Symmetrie adjustment"
and
"Diaphragm adjustment"
Sirona Dental Systems GmbH
(see
B
A
panorama
panorama
64 95 142 D3632
D3632.031.01.02.02 03.2016