13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
264
Diaphragm adjustment
13.2.2.2
✔
The 3D calibration block
1. In the structure tree, under
adjustment"
2. Establish readiness for exposure [ → 209].
3. Create an exposure (85 kV / 0.34 mAs) [ → 209].
A message is displayed in the action window to indicate whether
adjustment has been successful and the values can be applied.
4. Repeat the procedure starting at Step 2 until adjustment is
successful.
5. When adjustment is OK, save the values. [ → 209]
A message is displayed in the action window to indicate that the
values have been successfully saved.
The status indicator that prefixes the element
adjustment"
6. Continue with the next stage of the calibration procedure. [ → 265]
Installation Manual ORTHOPHOS SL
is not
in the bite block holder on the unit.
"3D"
, click on the
element (S030.10).
is checked.
Sirona Dental Systems GmbH
"Diaphragm
"Diaphragm
64 95 142 D3632
D3632.031.01.02.02 03.2016