2S021: Test Step 4 - Sirona ORTHOPHOS SL 2D Installation Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

11 Commissioning, demo mode and repacking
11.12 Service routines for commissioning and repacking
162
S021: Test step 4
11.12.5.2
Initiate the packing position for the sensor unit
1. Call service routine S021.4 [ → 133].
 An inactive progress indicator in selection field 1 and the
message FFFF in selection field 2 signal that the system is ready
to move the sensor unit.
2. Touch the T key (A).
 The sensor unit is moved to the packing position.
During this time, a progress indicator is displayed.
 When the packing position is reached, "0000" is displayed in
selection field 2.
3. Exit the service routine [ → 136].
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016