7 Installation: Remote control
7.7 Connecting a door contact switch
80
7.7
Connecting a door contact switch
X108
DX42
A
Door contact switch
PIN
1
2
3
4
5
6
➢ Connect the door contact switch between terminal X108 pin 1 (board
DX42) and control cable L117 pin 1 (BU).
Installation Manual ORTHOPHOS SL
Pin
X108
1
2
3
4
5
6
Code
WHGN
GN
BN
WHBN
BU/WHOG
WHBU/OG
Sirona Dental Systems GmbH
A
WHGN
GN
BN
WHBN
BU/WHOG
WHBU/OG
Appearance
White/green
Green
Brown
White/brown
Blue / white/orange
White/blue / orange
64 95 142 D3632