Electrical connection

8.1

Connecting the control cables

CAUTION
Potentially lethal shock hazard!
For PCs which are connected to an X-ray unit and are in the same room
as the unit, an additional protective ground wire is required (4mm
cable lug 4 - 6 DIN 46234 CU) in accordance with IEC 60601-1 (see also
"Installation Requirements ORTHOPHOS SL"
SIDEXIS
A
1. Use the media converter to connect the PC to the SC:SC socket on
the unit (see Operating Instructions of the media converter).
IMPORTANT
Use a duplex patch cable (L25), 1:1 connection, SC/SC 50/125 µm.
2. Install the media converter at a suitable location using fastening
screws or the Velcro strap supplied for this purpose. Do
media converter to the X-ray unit.
Installation version without remote control
➢ Connect the coiled cable of the release button to socket X103 of the
unit.
Installation version with remote control
1. Use cable L117 to connect the remote control to socket X103 on the
unit.
2. Secure the connector on socket X103.
Kat.5
RJ45
0
0 0 0 0 0
0 0
B
C
X108
X101
D
8 Electrical connection
8.1 Connecting the control cables
2
with
(REF 64 95 183)).
L25
SC1:1
L117
X103
Wagoklemme
not
attach the
81