11 Commissioning, demo mode and repacking
11.12 Service routines for commissioning and repacking
U: Adjusting and saving the upper value
L: Adjusting and saving the lower value
Checking the adjustment
158
1. Insert the occlusal bite block into the bite block holder and deflect the
bite block plate all the way up (up to the stop).
 "U" (for upper value) is displayed in selection field 1.
 The Memory key lights up.
2. To save the upper value "U", first press the Memory key (the R key
lights up) and then the R key.
 The value for the upper value "U" is accepted.
1. Deflect the bite block plate all the way down (up to the stop).
 "L" (for lower value) is displayed in selection field 1.
 The Memory key lights up.
2. To save the lower value "L", first press the Memory key (the R key
lights up) and then the R key.
 The value for the lower value "L" is accepted.
 The adjustment of the AD values is now finished.
1. Check the adjustment via service routine S018.7 [ → 153]. To do this,
move the bite block plate and check whether the adjustment range is
between -17° and +22°.
2. Exit the service routine.
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632