13 Unit adjustment and calibration
13.1 General information on unit adjustment and calibration
D
214
NOTICE
For correct calibration, the geometry phantom must be aligned
horizontally and vertically using a spirit level after insertion in the bite
block holder.
1. Insert the geometry phantom (A) into the pan bite block holder (B) on
the unit and align the geometry phantom correctly with the screw (C).
2. Attach the clip (D) for better stability.
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016