Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016
13.2.1.2.2
Manual adjustment: Pan - Diaphragm
The manual adjustment procedure is similar to the one for automatic
adjustment. The only difference is that the default adjustment values
calculated automatically in the
overwritten by adjustment values calculated manually.
1. Start the
described in the chapter entitled "Automatic adjustment: Pan
diaphragm (standard) [ → 221]".
2. Once you have taken the exposure, measure distances Sx and Sy
with the toolbar measuring ruler [ → 197].
Tip: For better measurement, you can also edit [ → 197] the exposure
using the toolbar functions.
13.2 Adjustment and calibration via the calibration menu
"Diaphragm adjustment"
"Diaphragm adjustment"
13 Unit adjustment and calibration
menu are
adjustment procedure as
225