Connecting the X-ray warning lamp

You can activate an X-ray warning lamp via the remote control.
NOTICE
A maximum load of 50W is permissible. No additional circuit may be
connected.
Use a 3-wire cable (1.5mm²) to connect the X-ray warning lamp.
1. Guide the cable (A) for connecting the X-ray warning lamp into the
chassis (concealed or surface installation).
2. Attach the cable to the still exposed strain relief.
A
C
X-ray warning lamp
3. Connect the current-carrying cables and the ground wire to terminal
X105 (board DX42) as shown in the connection diagram.
4. Secure the current-carrying cables with a cable tie (B) to prevent
them from slipping out.
5. Lay the cable (A) parallel to control cable L117 and close the housing.
X105
DX42
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
X108
X105
K100
DX42
64 95 142 D3632
D3632.031.01.02.02 03.2016