Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
Only for Germany
64 95 142 D3632
D3632.031.01.02.02 03.2016
13.2
Adjustment and calibration via the calibration
menu
IMPORTANT
After every adjustment and calibration of the unit, the reference values
for the constancy measurement must be recalculated and entered in the
"Test results" form, "Reference value" column (see technical document
"ORTHOPHOS SL 2D/3D Acceptance test DIN 6868 Part 151 / Part
161"
(REF 65 61 059).
13.2.1
2D adjustment/calibration
Pan sensor adjustment
13.2.1.1
13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
215