10 Installation: Ceph arm
10.3 Installation
C
4x
D
94
10.3.2

Ceph arm mounted on right-hand side

Preparing the unit
1. Loosen the screws (A).
A
2. CAUTION! Do not yank out the ambient light cable!
Remove the cover (B).
Tip: You can even leave the cover (B) on the unit and push it upwards
a little.
3. Remove the cover (C) by pulling it upwards.
B
4. Detach the cable fastening.
5. Unpack the cables.
6. Roll the cables up and fasten the cable roll above the bearing bolts
with adhesive tape.
7. Screw the four bearing bolts (D) into the unit.