Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
A
64 95 142 D3632
D3632.031.01.02.02 03.2016
IMPORTANT
After switch-on, the unit requires approx. 1 min. of warm-up time.
At the same time, the self-adjustment routine for the mechanics and the
electronics of the unit is performed. If a key is pressed during the self-
adjustment routine, an error message will display on the control panel.
1. Turn the main switch (A) to position I.
2. Wait for approx. 1 minute.
B
 The radiation indicator X-ray (C) lights up for approx. one second
 For a number of seconds the height adjustment keys light up
 After approx. 2 seconds, the green LED B in the upper part of the
C
 The start screen is displayed on the touchscreen for several
 The program selection is then displayed on the touchscreen.
 The forehead support and temple supports are completely open.
3. Check whether the patient symbols on the touchscreen can be
selected in exactly the right position.
If problems occur during selection, adjust the touchscreen [ → 272].
11 Commissioning, demo mode and repacking
as a functional check.
blue, then quickly change to green.
control panel lights up. This LED remains lit as long as the unit is
on.
seconds.
11.5 Switching the unit on
107