13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
236
A
Adjustment OK
B
Adjustment not OK
4. Repeat the procedure starting at Step 2 until the values in the menu's
text boxes are equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT: If automatic adjustment fails to produce values equal
to zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 237].
5. If adjustment is OK (the adjustment values in the text boxes are equal
to "zero"), save the values. [ → 209]
 The status indicator that prefixes the
primary diaphragm"
6. Remove the ceph test phantom from the sensor unit cover.
7. Continue with the next stage of the adjustment procedure. [ → 238]
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
● The vertical pin must be
horizontally centered in the
exposed image section. A slight
vertical offset of the grid is
permissible.
● A uniform white border
surrounding the image on all sides
must be visible.
● Distance S3 must be approx. 60
mm.
IMPORTANT: If S3is >70mm, please
contact the Sirona Customer Service
Center:
"Adjustment of Ceph
element is checked.
D3632.031.01.02.02 03.2016
64 95 142 D3632