Sirona ORTHOPHOS SL 2D Installation Manual page 232

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
232
6. Repeat the procedure from point 5 for all other kV/mA combinations.
 After the last exposure has been taken, a message in the action
window indicates whether calibration has been successful and
the status indicator at the start of the element
(DCS)"
in the structure tree responds accordingly:
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
7. Continue with the next stage of the adjustment procedure [ → 233].
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
"Sensor calibration
64 95 142 D3632
D3632.031.01.02.02 03.2016

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 2d / cephOrthophos sl 3dOrthophos sl 3d / ceph

Table of Contents