13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
244
3. Overwrite the adjustment values for S1, S2 and S3 in the text boxes
"Adjustment of Ceph secondary diaphragm"
of the
measured values. For information about the shifting direction (input
of +/- sign in the menu), see the section entitled "Shifting direction of
the exposed image area [ → 207]".
NOTICE! Use points as decimal separators!
4. Continue with the adjustment procedure as described in the chapter
entitled "Automatic adjustment: Ceph secondary diaphragm
(standard) [ → 238]"