Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016
13.2 Adjustment and calibration via the calibration menu
6. Take an exposure (85 kV / 0.34 mA s) [ → 209].
A message is displayed in the action window to indicate whether
adjustment has been successful and the status indicator in front
"8x8"
of the
element in the structure tree responds accordingly:
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
7.
If volume "11x10" is available:
"Next"
Click on the
button or select the volume
structure tree (S030.13).
"11x10"
The
volume is selected.
8. Establish [ → 209] receptivity.
9. Take an exposure (85 kV / 0.34 mA s) [ → 209].
A message is displayed in the action window to indicate whether
adjustment has been successful and the status indicator in front
"11x10"
of the
element in the structure tree responds
accordingly:
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
10. Continue with the next stage of the calibration procedure. [ → 268]
13 Unit adjustment and calibration
"11x10"
in the
267