5 Installation: X-ray unit
5.4 Installing the floor stand (option 2)
O
O
O
I
60
6. Remove the styrofoam packaging and rotate the X-ray tube assembly
counterclockwise to the front side of the unit.
7. Loosen screws (P) and adjusting nuts (O) again slightly.
8. Level the unit in both directions with the help of the spirit level by
adjusting the adjusting nuts (O):
Align the stand by placing the spirit level on the side and rear of the
stand (I).
Check the ring in both directions using the spirit level (J).
J
IMPORTANT
Be sure to tighten all adjusting nuts equally (to the same torque) after
leveling.
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016