Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016
Adjustment of the ceph primary diaphragm
13.2.1.5
13.2.1.5.1
Automatic adjustment: Ceph - Primary diaphragm (standard)
NOTICE
Phantom for Ceph exposure
Risk of damage to unit
It is essential that the needle phantom is removed from the bite block
holder on the unit before a ceph exposure is taken; otherwise, the
sensor could collide with the phantom.
✔
The needle phantom is
✔
The ceph test phantom connects to the cover of the sensor
unit [ → 212].
✔
The ceph sensor is plugged into the ceph slot on the unit.
✔
The ear plug holders on the ceph arm have been moved completely
apart and swung out of the beam direction (ap).
1. In the structure tree, under
primary diaphragm"
The
in the action area.
2. Establish [ → 209] receptivity.
13.2 Adjustment and calibration via the calibration menu
not
in the bite block holder on the unit.
"Ceph"
element (S010.3).
"Adjustment of Ceph primary diaphragm"
13 Unit adjustment and calibration
"Adjustment of Ceph
, click on the
menu is displayed