13 Unit adjustment and calibration
13.3 Adjusting the touchscreen via the web service
272
13.3

Adjusting the touchscreen via the web service

The unit and computer are turned on.
The unit is logged into the network as an X-ray component.
1. Start an Internet browser such as Internet Explorer or Firefox.
2. In the address line, enter "https://" and the IP address of your unit.
Example: https://192.168.15.240
Tip: The IP address of your unit can be found on the info screen.
3. Confirm your input with the "Enter" key.
 The Sirona Web service is loaded.
 The sub menus of the
tree.
4. Click on the
 The service routine for touchscreen calibration is opened.
 You can abort the adjustment procedure at any time by clicking
on the
5. Click on the
Installation Manual ORTHOPHOS SL
"User"
menu are displayed in the structure
"Easypad Touchcalibration"
"Cancel"
button in the Web browser.
"Display Easypad Touchcalibration Screen"
Sirona Dental Systems GmbH
button.
button.
64 95 142 D3632
D3632.031.01.02.02 03.2016