Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
"Diaphragm adjustment"
(Pan - adjustment)
64 95 142 D3632
D3632.031.01.02.02 03.2016
13.1 General information on unit adjustment and calibration
Diaphragm adjustment can usually be performed directly via precision
adjustment. Only in exceptional cases, e.g. if the exposed image area is
completely outside the image section in an image acquired with the
adjustment"
setting (C), is it necessary to perform coarse adjustment
prior to precision adjustment (D).
On the exposure with coarse adjustment (D), the exposed area is just
barely visible at the right margin of the image field. Even in this extreme
case, an adjustment would still be possible.
13 Unit adjustment and calibration
"Fine
205