Sirona ORTHOPHOS SL 2D Installation Manual page 234

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
234
3. Create an exposure (64 kV / 16 mA; 6.1 s) [ → 209].
 The adjustment values S1, S2 S3 and S4 are automatically
determined from the exposure and entered in the text boxes of
"Adjustment of Ceph primary diaphragm"
the
 The exposure is displayed in the exposure window.
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
menu.
64 95 142 D3632
D3632.031.01.02.02 03.2016