13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
234
3. Create an exposure (64 kV / 16 mA; 6.1 s) [ → 209].
The adjustment values S1, S2 S3 and S4 are automatically
determined from the exposure and entered in the text boxes of
"Adjustment of Ceph primary diaphragm"
the
The exposure is displayed in the exposure window.
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
menu.
64 95 142 D3632
D3632.031.01.02.02 03.2016