Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
1x M10
1x
R
Q
64 95 142 D3632
D3632.031.01.02.02 03.2016
Setting up and leveling the unit
1. Remove the transport safety device (see section "Remove the
transport safety device [ → 63]").
2. Insert screw (P) through the washer and then from the rear through
the support and into the stand.
P
1x Ø 10,5
1x M10x50
3. Fit the profile clamp (Q) onto the screw (P) from the other (front) side
and screw the nut (R) onto the screw (P).
4. Tighten adjusting nuts (O) and screw (P) firmly.
IMPORTANT
Observe the required movement range of the unit during installation.
5. Set up the unit including the central styrofoam packaging and position
it in its ultimate location.
5 Installation: X-ray unit
5.4 Installing the floor stand (option 2)
Q
R
P
59