Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016
Pan aperture adjustment
13.2.1.2
13.2.1.2.1
Automatic adjustment: Pan diaphragm (standard)
IMPORTANT
If a message window indicating that the diaphragm is tilted appears in
the menu during
mechanically maladjusted in the vertical axis. In this case, adjust the
diaphragm [ → 237] and repeat after the exposure.
IMPORTANT
If a message window indicating that another X-ray has to be taken
appears in the menu during
procedure starting at Step 2, even if the values in the menu's text boxes
are equal to zero. This additional exposure ensures that the diaphragm
gap width offset is also adjusted correctly.
13.2 Adjustment and calibration via the calibration menu
"Diaphragm adjustment"
, the primary diaphragm is
"Diaphragm adjustment"
13 Unit adjustment and calibration
, repeat the
221