13 Unit adjustment and calibration
13.1 General information on unit adjustment and calibration
212

Adjustment phantom for adjustment of the cephalometer

13.1.6.2
"Adjustment of Ceph primary diaphragm"
For
main beam direction"
(A) must be inserted into the sensor unit cover.
"Adjustment of Ceph secondary diaphragm"
For the
the ceph adjustment phantom must be removed from the sensor unit
cover once again.
Installation Manual ORTHOPHOS SL
A
ceph adjustments, the ceph adjustment phantom
Sirona Dental Systems GmbH
"Adjustment of Ceph
and
ceph adjustment,
64 95 142 D3632