13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
262
13.2.2

3D adjustment/calibration

IMPORTANT
For the 3D adjustment and calibration, the values in the input fields do
not have to be equal to zero.
Confirmation of a successful adjustment and calibration is the message
in the calibration menu action window; e.g. "Adjustment successful",
"Values can be adopted", etc.

Sensor adjustment

13.2.2.1
1. Insert the 3D calibration block into the bite block holder on the
unit. [ → 213]
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632