Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016
Adjustment of the ceph main X-ray beam
13.2.1.8
13.2.1.8.1
Automatic adjustment: Ceph main X-ray beam direction (standard)
NOTICE
Risk of damage to unit
It is essential that the needle phantom is removed from the pan bite
block holder on the unit before a ceph exposure is taken; otherwise, the
sensor unit could collide with the phantom.
✔
The ceph test phantom connects to the unit's sensor unit [ → 212].
✔
The ear plug holders on the ceph arm have been swung out of the
beam direction (ap).
1. In the structure tree, under
main beam direction"
The
displayed in the action area.
2. Establish [ → 209] receptivity.
13.2 Adjustment and calibration via the calibration menu
"Ceph"
element (S010.6).
"Adjustment of Ceph main beam direction"
13 Unit adjustment and calibration
"Adjustment of Ceph
, click on the
menu is
253