Adjustment Of The Ceph Main X-Ray Beam - Sirona ORTHOPHOS SL 2D Installation Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016

Adjustment of the ceph main X-ray beam

13.2.1.8
13.2.1.8.1
Automatic adjustment: Ceph main X-ray beam direction (standard)
NOTICE
Risk of damage to unit
It is essential that the needle phantom is removed from the pan bite
block holder on the unit before a ceph exposure is taken; otherwise, the
sensor unit could collide with the phantom.
The ceph test phantom connects to the unit's sensor unit [ → 212].
The ear plug holders on the ceph arm have been swung out of the
beam direction (ap).
1. In the structure tree, under
main beam direction"
 The
displayed in the action area.
2. Establish [ → 209] receptivity.
13.2 Adjustment and calibration via the calibration menu
"Ceph"
element (S010.6).
"Adjustment of Ceph main beam direction"
13 Unit adjustment and calibration
"Adjustment of Ceph
, click on the
menu is
253