2S032: Test Step 50 - Sirona ORTHOPHOS SL 2D Installation Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

11 Commissioning, demo mode and repacking
11.12 Service routines for commissioning and repacking
164
S032: Test step 50
11.12.6.2
Self-test of the FP image data path
1. Call service routine S032.50 [ → 133].
 Once the service routine has been selected, "FP test" is
displayed in selection field 1.
2. Touch the T key (B).
 The FP image data path is tested.
 If the image data path is OK, "Success" is displayed in the
selection field 1.
3. Exit the service routine [ → 136].
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016