Performing A Mechanical Function Test; Starting The Test Cycle - Sirona ORTHOPHOS SL 2D Installation Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
T
Ready for exposure
64 95 142 D3632
D3632.031.01.02.02 03.2016
11.6

Performing a mechanical function test

The mechanical function of the rotation unit is checked via a test cycle.
The test cycle is executed without radiation. The test cycle is used to
check that the unit is functioning correctly and to ensure that a complete,
uninterrupted cycle is possible. The rotating unit stops automatically if the
resistance increases.
If any untypical noises occur during the test cycle or the rotation unit stops
due to increased resistance, the assembly of the sensor must be checked
again.
11.6.1

Starting the test cycle

The test cycle is executed without radiation. The test cycle is used to
check that the unit is functioning correctly and to ensure that a complete,
uninterrupted cycle is possible. The rotating unit stops automatically if the
resistance increases.
The unit is in its starting position.
1. Press the T key.
 The program enters test cycle mode. On the touchscreen, the
display of the kV/mA value, the exposure time and the patient
symbols is hidden. Two test cycle symbols appear.
2. Press the release button.
 The test cycle is started.
3. Wait until the test cycle has been completed.
4. Press the T key again.
 The program exits test cycle mode.
11 Commissioning, demo mode and repacking
11.6 Performing a mechanical function test
109