11 Commissioning, demo mode and repacking
11.12 Service routines for commissioning and repacking
Adjusting and saving the PWM index
values
160
1. Call service routine S018.10.
 The PWM index for upward travel "U" (up) is displayed in
selection field 1.
 The PWM index for downward travel "D" (down) is displayed in
selection field 2.
 The display of patient symbol key 1 (G) indicates whether or not
the occlusal bite block is connected (see table).
2. Select the desired index values using the arrow keys in selection
fields 1 and 2 (see table).
3. To save the index values, first press the Memory key (the R key lights
up) and then the R key.
4. Exit the service routine.
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016