Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016
Adjustment of the ceph secondary diaphragm (QuickShot)
13.2.1.7
13.2.1.7.1
Automatic adjustment: Ceph secondary diaphragm (QuickShot)
(standard)
NOTICE
Phantom for Ceph exposure
Risk of damage to unit
It is essential that the needle phantom is removed from the bite block
holder on the unit before a ceph exposure is taken; otherwise, the
sensor could collide with the phantom.
✔
The pan needle phantom is
unit.
✔
The ceph test phantom does
unit.
1. In the structure tree, under
secondary diaphragm (Quickshot)"
The
menu is displayed in the action area.
2. Establish [ → 209] receptivity.
13.2 Adjustment and calibration via the calibration menu
not
in the pan bite block holder on the
not
connect to the cover of the sensor
"Ceph"
"Adjustment of Ceph secondary diaphragm (Quickshot)"
13 Unit adjustment and calibration
"Adjustment of Ceph
, click on the
element (S010.8).
247